Emerson Process Management ROSEMOUNT 5300 Instrukcja Użytkownika Strona 192

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 324
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 191
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
A-8
Remote Housing
Measuring Range
This table shows the maximum recommended measuring range with Remote
Housing for different RH lengths, installation types, Dielectric Constants, and
probe types.
Table A-2. Remote Housing
measuring range
Dielectric
Constant
Rigid Single
8 mm
Rigid Single
13 mm
Flexible Single Coaxial Rigid Twin Flexible Twin
1 m
Remote
Housing
Chamber / pipe
installations 4 in.
(100 mm)
1.4
2
80
1.4
2
80
4 ft (1.25 m)
10 ft (3 m)
(1)
10 ft (3 m)
4 ft (1.25 m)
4 ft (1.25 m)
10 ft (3 m)
(1)
15 ft (4.5 m)
(1)
15 ft (4.5 m)
(1)
15 ft (4.5 m)
(1)
4 ft (1.25 m)
4 ft (1.25 m)
10 ft (3 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
4 ft (1.25 m)
4 ft (1.25 m)
159 ft (48.5 m)
(1)
19 ft
(6 m)
10 ft (3 m)
(1)
10 ft (3 m)
(1)
10 ft (3 m)
(1)
4 ft (1.25 m)
4 ft (1.25 m)
10 ft (3 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
4 ft (1.25 m)
98 ft (30 m)
(1)
159 ft (48.5 m)
(1)
Tank installations
2 m
Remote
Housing
Chamber / pipe
installations 4 in.
(100 mm)
1.4
2
80
1.4
2
80
9 ft (2.75 m)
10 ft (3 m)
(1)
10 ft (3 m)
9 ft (2.75 m)
9 ft (2.75 m)
10 ft (3 m)
(1)
15 ft (4.5 m)
(1)
15 ft (4.5 m)
(1)
15 ft (4.5 m)
9 ft (2.75 m)
9 ft (2.75 m)
10 ft (3 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
9 ft (2.75 m)
9 ft (2.75 m)
154 ft (47 m)
(1)
19 ft
(6 m)
10 ft (3 m)
(1)
10 ft (3 m)
(1)
10 ft (3 m)
(1)
9 ft (2.75 m)
9 ft (2.75 m)
10 ft (3 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
9 ft (2.75 m)
98 ft (30 m)
(1)
154 ft (47 m)
(1)
Tank installations
3 m
Remote
Housing
Chamber / pipe
installations 4 in.
(100 mm)
1.4
2
80
1.4
2
80
10 ft (3 m)
15 ft (4.5 m)
15 ft (4.5 m)
15 ft (4.5 m)
14 ft (4.25 m)
14 ft (4.25 m)
15 ft (4.5 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
14 ft (4.25 m)
14 ft (4.25 m)
149 ft (45.5 m)
(1)
19 ft
(6 m)
10 ft (3 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
14 ft (4.25 m)
98 ft (30 m)
(1)
149 ft (45.5 m)
(1)
Tank installations
(1) Accuracy may be affected up to + 1.2 in. (30 mm).
(2) Required chamber/pipe size is 3 or 4 in. (75 -100 mm).
Przeglądanie stron 191
1 2 ... 187 188 189 190 191 192 193 194 195 196 197 ... 323 324

Komentarze do niniejszej Instrukcji

Brak uwag