Emerson Process Management Series 100 Gas Analyzers HYDROS 100 Instrukcja Użytkownika Strona 202

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 225
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 201
CHECKING AND REPLACING AN ELECTROCHEMICAL OXYGEN SENSOR
25 - 6
ETC00781(4) Series 100 e 02/2004
25.2.2 Removing the Sensor
P Take the consumed sensor out of the fitting.
P Take off the stopper from new sensor and place the new sensor into the fitting, so
that the name plate is at the top of the sensor.
P Close the consumed sensor with the stopper and send it back to our factory.
25.2.3 Reinstalling the Sensor
a) Oxygen Measurement without IR - channel
P Put the fitting with the (new) sensor onto the support, move to the stop and screw with
the two fastening screws (phillips screws, see Fig. 25-3)
P Fix the sensor with a cable tie to the support (see Fig. 25-3).
P Connect the connector for the sensor to “P2” of circuit board “OXS” (see Fig. 25-5 / chapter 18.).
P Insert the complete support (see Fig. 25-3) into the analyzer and fix it with the two
allen screws (fastening screws, see Fig. 25-2).
P Reconnect all gas lines to the fittings (see Fig. 25-2 and Fig. 25-3).
Do not interchange gas inlets and gas outlets.
P Reconnect all electrical connections between OXS and BKS (X5 and X7, see chapter 18.).
P Perform a leakage test (see chapter 22.) and set the sensor (see chapter 25.2.4).
b) Combined IR / Oxygen Measurement
P Put the fitting with the (new) sensor onto the support, move to the stop and fix it with
the two fastening screws (see Fig. 25-4)
P Connect the connector for the sensor to “P2” of circuit board “OXS” (see Fig. 25-5 / chapter 18.).
P Perform a leakage test (see chapter 22.) and set the sensor (see chapter 25.2.4).
REPLACE / REINSTALL THE SENSOR
Przeglądanie stron 201
1 2 ... 197 198 199 200 201 202 203 204 205 206 207 ... 224 225

Komentarze do niniejszej Instrukcji

Brak uwag